Description
The cryocompressor APD HC-8C is installed in magnets Philips NT, F2000. This cryocompressor works complete with cold heads SC-8C, SC-R8. The HC-8C adsorber is subject to scheduled replacement every 20 thousand hours of operation (28 months) and with each replacement of the cold head to prevent contamination of the cryogenic system